Module No. 3515

Calibrating Temperature Readings

The watch’s temperature sensor is adjusted at the factory and normally does not require calibration. However, you can calibrate the displayed value if you notice major errors in readings.

Important!

Check to make sure that the value you use for calibration produces correct temperature readings.

Before calibrating the temperature reading, remove the watch from your wrist and leave it in the area where you plan to measure temperature for about 20 or 30 minutes to allow the case temperature to become the same as the air temperature.

  1. Display the current time screen.

  2. Use (A) and (D) to select [BAROMETER].

  3. 3515_BARO
  4. Press (C).

  5. This displays the current barometric pressure and temperature.

  6. Hold down (A) for at least two seconds.

  7. This displays the barometer/temperature measurement setting screen.

    3515_BAR_TEM_set_caps
  8. Use (A) and (D) to move the pointer to [THERMOMETER CALIBRATION].

  9. Press (C).

  10. Use (A) and (D) to move the pointer to [MANUAL].

  11. Selecting [OFFSET OFF] returns barometric pressure calibration to its factory default state.

  12. Press (C).

  13. This displays the manual correction screen.

  14. Use (A) and (D) to input the current temperature.

  15. 3515_074_BAR_TEM
  16. Press (C) to complete the setting operation.

  17. After the setting operation is complete, a check mark appears for a few seconds, and then the barometric pressure/temperature setting screen re-appears.

  18. Press (E) to exit the setting screen.

Back

Next